>> Infusion Pump - เครื่องให้สารละลายทางหลอดเลือดดํา
>>  
3
AKAS SYRU 4004 ver 4

Power Supply: 220 ± 10% V AC, 50 Hz

Battery: 12V 800mAh NiMH battery (Optional 11.1 V LiPO Battery)

Battery Backup: 8 hours @ 5 ml/hr flow rate on 50 ml syringe

Power Consumption: 7 VA

Display: LCD & LED Display

Applicable Syringe: 2 ml, 5 ml, 10 ml, 20 ml, 30/35 ml, 50/60 ml syringes from any brand. Custom programmable for any brand syringe also available.

Automatic Syringe Size Detector: Available

Flow Rate Range:

  • Mains:

    • 0.01-350 ml/hr for 2 ml syringe
    • 0.1-750 ml/hr for 5 ml syringe
    • 0.1-1000 ml/hr for 10 ml syringe
    • 0.1-1800 ml/hr for 20 ml syringe
    • 0.1-2500 ml/hr for 30 ml syringe
    • 0.1-3000 ml/hr for 50 ml syringe
  • Battery:

    • 0.01-180 ml/hr for 2 ml syringe
    • 0.1-350 ml/hr for 5 ml syringe
    • 0.1-500 ml/hr for 10 ml syringe
    • 0.1-900 ml/hr for 20 ml syringe
    • 0.1-1200 ml/hr for 30 ml syringe
    • 0.1-2000 ml/hr for 50 ml syringe

Infused Volume: 0.1 - 9999.9 ml

Target Volume: 0.1 to 9999.9 ml

Elapsed Time: Displays elapsed time from 00:01 to 199:59 hours

Ramp Up/Taper Down Dosage Mode: Available

Sequential Mode Volume: Available

KVO (Keep Vein Open): 0.1 ml/hr to 9.9 ml/hr in 0.1 ml increments

Bolus Purge: Bolus & Auto Bolus

Bolus Volume: 0.1 ml to 10 ml in 0.1 ml increments

Occlusion Pressure: 10 steps selection (OFF to 9 Levels)

Anti-Bolus Function: Post occlusion bolus reduction function to avoid accidental bolus delivery after occlusion incident

Event Log with RTC: Available for 5000 events

Alarms: Target volume reached, target time reached, end of syringe, PCA probe disconnected, occlusion pressure limit reached, occlusion pressure pre-alarm, battery discharged, syringe holder disturbed during infusion, delivery line disconnected alarm, internal mechanism fault, mains to battery changed, flow rate reduced to battery range, mains voltage over the limit, mains voltage below 215 reduce flow rate ≤2000 ml/hr, priming not done, occlusion sensor failed, low patient back pressure, maintain bed level equal to pump

Accuracy: Mechanical Accuracy ±1%; Including Syringe ±2%

Safety Testing: EN 60601-1, EN 60601-1-2, EN 60601-1-6, EN 60601-1-8, EN 60601-2-24

Certification: European CE, ISO 13485:2016, CDSCO registered Indian Medical Device Manufacturer

Space Occupied: 2.2 Liters

Weight & Dimension: 1.9 kg; 170 x 96 x 133 mm (L x B x H)

Inline Changing of Delivery Rate: Possible to change the infusion rate without stopping the infusion

Last Data ON/OFF: Saves the last infusion data even when power OFF & ON

Alarm Volume: Adjustable 5 levels

Pause Infusion Facility: Standby option from 1 min to 24 hours, user adjustable

Battery Remaining Time: Viewable under settings

Adjustable Near to End Time: From 3 min to 30 min, flow rate dependent only

AMC Reminder: Available

Keypad Locking: Available, manual & automatic (ON/OFF option)

Pressure Display: Inline pressure display available in PSI value (ON/OFF option)

Alarm LED Brightness: Adjustable 5 levels

Optional Pole Clamp: Multifunction mounting pole clamp for horizontal/vertical pole mounting

Unipower Supply (Optional): 5-in-1 power cord supply for stacked pumps

Docking Stand (Optional): Stack 4 pumps & mountable on poles with unipower supply cord

Operating & Storage Conditions:

  • Temperature: -10 to 50°C
  • Humidity: 15 to 90%
  • Pressure: 50 to 106 kPa

Datasheet